1.1 ,This specification covers polyethylene plastics (as defined in Terminology D883) intended for use in medical device applications involving human tissue contact devices, short-term indwellings of 30 days or less, and fluid transfer devices. The biocompatibility of these materials as a class has not been established. Biocompatibility tests must be conducted on the final product.