1.1 ,This performance specification establishes performance requirements for the short-term utilization of a single-use, balloon-retention catheter, French sizes 12 through 26 inclusive, used by the medical professions for providing a means of bladder drainage by means of the urethra. The product is manufactured in various sizes and materials such as latex, silicone, rubber, and various polymers (as well as combinations of these) and is provided non-sterile for sterilization and sterile for single use only. Catheters whose surface has been chemically treated to effect biocompatibility or microbial properties may be tested to this specification.